TROUBLESHOOTING
7-26 Sysmex KX-21 Operator’s Manual -- Revised October 1998
Error Message Meaning of Message
"Calibration Error"
ERR CODE:XXXXX.X.X
*
Help
*
Calibration value is out of range.
If error persists,
>=5% difference from the last calib.
Calib. factor is out of 80%-120%.
3:Return
The set calibration value does not meet the
conditions:
• Difference from the last calibration must not
exceed 5%.
• Calibration value must be in the range of
80% - 120%.
Cause 1) Input error of target or calibration values
2) Instrument failure caused data to shift.
Action
1) Checking Entered Values
Press [3] key to return to the Calibration screen and check the
target or calibration value.
• In case of input error of target value in automatic calibration
Select "No" on the Update screen for calibration value and
execute Auto Calibration again from the beginning.
• In case of input error of calibration value in manual calibration
Select "No" on the Update screen for calibration value and enter
calibration value again.
2) Maintenance of Instrument
Check QC data, and if data is found to have shifted, a hardware
problem is conceivable. Referring to
Chapter
4: INSTRUMENT
MAINTENANCE AND SUPPLIES REPLACEMENT,
clean the
transducer, SRV, etc.
Analysis after error
occurrence
Not Ready until calibration value is fixed.