FUNCTIONAL DESCRIPTION
9-26 Sysmex KX-21 Operator’s Manual -- Revised October 1998
NOTE: Since CBC8 parameters only are output in the pre-diluted mode, flag
output is limited to a PLT parameter.
1) 1A High frequency for LD
PLT, PDW, MPV, and P-LCR parameters are flagged with PL.
Platelet Histogram Error - PL (1A)
211
12.5
9.3
-22.6
PLT
[×10
3
/µL]
[fL]
[fL]
[%]
302010 [fL]
PLT
PDW
MPV
P-LCR
PL
PL
PL
PL
Figure 9-6-20: Platelet Histogram Error - PL (1A)
2) 1B High LD with 20%-frequency distribution width error
PLT parameter is flagged with PL. Particle distribution width (PDW) is flagged
with DW and its data are not output.
MPV and P-LCR are flagged with PL and their data are not output.
3) 1C High LD with two or more peaks
PLT parameter is flagged with PL. Particle distribution width (PDW) is flagged
with MP and its data are not output.
MPV and P-LCR are flagged with PL and their data are not output.
4) 2A High UD
PLT, PDW, MPV, and P-LCR parameters are flagged with PU.
5) 2B High UD with 20%-frequency particle distribution width error
PLT parameter is flagged with PU. Particle distribution width (PDW) is flagged
with DW and its data are not output.
MPV and P-LCR are flagged with PU and their data are not output.
Platelet Histogram Error - PU (2B)
171
---.-
---.-
---.-
PLT
[×10
3
/µL]
[fL]
[fL]
[%]
302010 [fL]
PLT
PDW
MPV
P-LCR
PU
DW
PU
PU
Figure 9-6-21: Platelet Histogram Error - PU (2B)