EasyManua.ls Logo

SYSMEX KX-21 - Page 244

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FUNCTIONAL DESCRIPTION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 9-27
6) 2C High UD with two or more peaks
PLT parameter is flagged with PU. Particle distribution width (PDW) is flagged
with MP and its data are not output.
MPV and P-LCR are flagged with PU and their data are not output.
7) 3 20%-frequency particle distribution width error
PLT is not flagged. PDW, MPV, and P-LCR are flagged with DW and their
data are not output.
Platelet Histogram Error - DW (3)
91
---.-
---.-
---.-
PLT
[×10
3
/µL]
[fL]
[fL]
[%]
302010 [fL]
PLT
PDW
MPV
P-LCR
DW
DW
DW
Figure 9-6-22: Platelet Histogram Error - DW (3)
8) 4 Two or more peaks
PLT is not flagged. Other parameters are flagged with MP and their data are not
output.
Platelet Histogram Error - MP (4)
155
---.-
---.-
---.-
PLT
[×10
3
/µL]
[fL]
[fL]
[%]
302010 [fL]
PLT
PDW
MPV
P-LCR
MP
MP
MP
Figure 9-6-23: Platelet Histogram Error - MP (4)

Related product manuals