FUNCTIONAL DESCRIPTION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 9-31
8.2 Front Interior
1) Detector Block
5) Built-in Printer
2) Sample Rotor Valve (SRV)
3) Rinse Cup
4) WBC/HGB Lyse
(STROMATOLYSER-WH)
Figure 9-8-2: Front Interior
1) Detector Block
Incorporates RBC transducer, WBC transducer, and HGB flow cell.
2) Sample Rotor Valve (SRV)
Makes volumetric measurement of aspirated blood.
3) Rinse Cup
Cleans the sample probe.
4) WBC/HGB Lyse (STROMATOLYSER-WH)
Reagent to measure WBC/HGB
5) Built-in Printer
Prints analysis data, error messages, etc.
Sysmex KX-21 Operator’s Manual -- Revised January 2000