EasyManua.ls Logo

SYSMEX KX-21 - Page 256

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT SETUP
Sysmex KX-21 Operator’s Manual -- Revised October 1998 10-3
(5) Press [ENTER] key. The current setting status is displayed on the System Setup
screen.
Units Type2
RDW RDW-SD
Parameter Naming LYM%
Language English
*
System Setup
*
Press [SELECT] to exit.
Figure 10-2-2: System Setup Screen
(6) Using [ ] or [ ] key, move the cursor to select the desired set parameter.
For System Setup, four parameters can be set.
(7) Using [ ] or [ ] key, select the content of settings. Each time you press the key, the
content of settings changes.
(8) Press [ENTER] key. The content of settings is set and the cursor moves to the next set
parameter.
NOTE: The settings can also be set when you press [ ] or [ ] key after
changing the content of settings.
1) Unit Setting
Sets the unit system for displaying analysis data. The unit can be selected from
the following six kinds:
Settings Unit System Settings Unit System
Type 1 For Japan Type 4 Dutch SI
Type 2 For General Export Type 5 Standard SI
Type 3 Canada SI Type 6 Hong Kong SI
Table 10-2-1: Unit Setting
For the detail of unit system, refer to the following table: