EasyManua.ls Logo

SYSMEX KX-21 - Page 270

SYSMEX KX-21
306 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
INSTRUMENT SETUP
Sysmex KX-21 Operator’s Manual -- Revised October 1998 10-17
(5) Press [ENTER] key. The current setting status is displayed on the Host Settings
screen.
Auto Output On
Output Format K-1000
Data Length 7bits
Baud Rate 2400
Stop Bit 2bit
Protocol Class A
Interval 2
RTS/CTS Ignore
Parity Check Even
*
Host Settings
*
Press [SELECT] to exit.
ConnectConnect Not UseNot Use
Figure 10-6-2: Host Settings Screen
(6) Using [
] or [ ] key, move the cursor to select the desired set parameter. For Host
Settings, eight parameters can be set.
(7) Using [ ] or [ ] key, select the condition. The condition changes each time you
press the key.
(8) Press [ENTER] key. The content of settings is set and the cursor moves to the next set
parameter.
NOTE: • The settings can also be set when you press [ ] or [ ] key after
changing the content of settings.
1) Connect Settings
Set whether to connect the host computer or not.
Settings Connection
Use Connect
Not Use Not Connect
Table 10-6-1: Host Output Settings