TECHNICAL INFORMATION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 B-5
(1) Analysis data format (121 characters)
Parameter No. of Characters Example
STX 1
(
02 H
)
Text distinction code I 1 "D"
Text distinction code ll 1 "1"
Sample distinction code 1 "U"
Year
(
Month or Da
y)
2 "97"
Month
(
or Da
y)
2 "09"
Da
y
(
or Year
)
2 "30"
Anal
y
sis information 1
(
O
)
Sample ID No. 12
(
OOOOOOOOOOOO
)
PDA information 6
(
OOOOOO
)
RDW select information 1 "S" or "C"
WBC [×10
3
/
µ
L] 5 XXX.XF
RBC [×10
6
/
µ
L] 5 XX.XXF
HGB [
g
/dL] 5 XXX.XF
HCT [%] 5 XXX.XF
MCV [fL] 5 XXX.XF
MCH [p
g
] 5 XXX.XF
MCHC [
g
/dL] 5 XXX.XF
PLT [×10
3
/
µ
L] 5 XXXXF
LYM%
(
W-SCR
)
[%] 5 XXX.XF
MXD%
(
W-MCR
)
[%] 5 XXX.XF
NEUT%
(
W-LCR
)
[%] 5 XXX.XF
LYM#
(
W-SCC
)
[×10
3
/
µ
L] 5 XXX.XF
MXD#
(
W-MCC
)
[×10
3
/
µ
L] 5 XXX.XF
NEUT#
(
W-LCC
)
[×10
3
/
µ
L] 5 XXX.XF
RDW-SD/CV [fL/%] 5 XXX.XF
PDW [fL] 5 XXX.XF
MPV [fL] 5 XXX.XF
P-LCR [%] 5 XXX.XF
ETX 1
(
03 H
)
Table B-1-4: Analysis Data Format