TECHNICAL INFORMATION
B-6 Sysmex KX-21 Operator’s Manual -- Revised October 1998
(2) QC data format (99 characters)
Parameter No. of Characters Example
STX 1 (02 H)
Text distinction code I 1 "D"
Text distinction code II 1 "1"
Sample distinction code 1 "C"
Data distinction code 1 "X" or "L"
Year (Month or Day) 2 "97"
Month (or Day) 2 "09"
Day (or Year) 2 "30"
Hour 2 "09"
Minute 2 "59"
QC File Number 1 "5"
Not used 1 "0"
RDW select information 1 "C"
WBC [×10
3
/µL] 4 XXX.X
LYM% (W-SCR) [%] 4 XXX.X
MXD% (W-MCR) [%] 4 XXX.X
NEUT% (W-LCR) [%] 4 XXX.X
LYM# (W-SCC) [×10
3
/µL] 4 XXX.X
MXD# (W-MCC) [×10
3
/µL] 4 XXX.X
NEUT# (W-LCC) [×10
3
/µL] 4 XXX.X
RBC [×10
6
/µL] 4 XX.XX
HGB [g/dL] 4 XXX.X
HCT [%] 4 XXX.X
MCV [fL] 4 XXX.X
MCH [pg] 4 XXX.X
MCHC [g/dL] 4 XXX.X
RDW-(CV/SD) [%/fL] 4 XXX.X
PLT [×10
3
/µL] 4 XXXX
PDW [fL] 4 XXX.X
MPV [fL] 4 XXX.X
P-LCR [%] 4 XXX.X
W-SMV [fL] 4 XXX.X
W-LMV [fL] 4 XXX.X
ETX 1 (03 H)
Table B-1-5: QC data format