TECHNICAL INFORMATION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 B-7
(3) Details of the analysis data output in K-1000 format
1) The order of output is from the top parameter to the bottom. The data is sent
from the upper line without performing zero suppression.
2) Because the decimal point is not sent, the decimal point specified in each
parameter has to be added and represented by numeric value at the host
computer.
3) The order of the date conforms to the format in the setting program in the setting
program.
4) Analysis information shows the following content of the analysis.
Manual No. input
WB mode analysis "0"
Pre-diluted mode analysis "5"
Table B-1-6: Analysis Information
5) The sample ID No. is 6-digit numerals, but 12 digits are output in this format.
The upper 6 digits are 0’s (zeros). Depending on the usage, "-" (2D H) may be
inserted in the numeral column. In this case, "-" is included in the 12 digits.
6) PDA information consists of the following 6 items.
Item No. of Characters
WBC histogram information 1
WBC histogram flag 1
RBC histogram information 1
RBC histogram flag 1
PLT histogram information 1
PLT histogram flag 1
Table B-1-7: PDA Information
Code Description
"0" Histogram is normal.
"1" Histogram is abnormal.
"2"
Manually discriminated
Table B-1-8: Histogram Information