TECHNICAL INFORMATION
Sysmex KX-21 Operator’s Manual -- Revised October 1998 B-9
7) RDW select information
It shows whether RDW-CV or RDW-SD is output.
"C": RDW-CV
"S": RDW-SD
NOTE: RDW-SD is selected at the time of shipping from the factory.
To output RDW-CV, refer to
Chapter 10, Section 2: System Setup
.
8) The numerical data configuration in the Analysis Data Format is as follows.
Upper digit Lower digit
Data Flag
Figure B-1-5: Numerical Data Configuration
Code Description
"0" Normal
"1" Abnormality judgment is "+."
"2" Abnormality judgment is "-."
"3" Out of linearity limit
"4" Reliability is low.
Table B-1-11: Details of Flag
9) If the data over-flows, it is indicated as follows:
*0003
10) In case of analysis error, or if the data lacks some data such as the calculation
parameter in pre-diluted mode, the data is transmitted as follows:
*0000