SAMPLE ANALYSIS
Sysmex KX-21 Operator’s Manual -- Revised October 1998 2-31
(2) When executing shutdown, set CELLCLEAN to the sample probe and press the start
switch in that status. While "Aspirating" is being displayed on the screen, keep
holding CELLCLEAN in the same status.
Start Switch
Figure 2-9-2: Pressing the Start Switch
Because CELLCLEAN is a strong alkaline detergent, take care not to
have it adhere to the skin or clothes. If the skin or clothes should
come in touch with it, flush it away using plenty of water. Otherwise,
it can damage the skin or clothes.
NOTE: Once shutdown is executed, it cannot be canceled until it is completed.
(3) When the buzzer sounds two times - beep, beep- , informing the completion of
aspiration, remove CELLCLEAN from the sample probe.
After that, shutdown is executed automatically.
*
Shutdown
*
Please wait.
Rinsing
Figure 2-9-3: Shutdown Execution Screen
CAUTION: To remove CELLCLEAN, lower it straight down. Take care not to bend
the sample probe.
WARNING