SAMPLE ANALYSIS
2-30 Sysmex KX-21 Operator’s Manual -- Revised October 1998
9. EXECUTION OF SHUTDOWN
When analyses of all samples are finished, execute shutdown before turning the power off.
By executing shutdown, the TD chamber and diluted sample line are cleaned. Execute
shutdown after the day's work is finished, and also when this instrument is used
continuously, shut it down at least once in 24 hours.
CAUTION: If the power is turned OFF before executing shutdown, failures could
occur, such as fluid dripping from the rinse cup and crystals forming
around the rinse cup.
9.1 Shutdown Procedure
(1) Press [SHUTDOWN] key in the Ready status. The Shutdown screen appears.
3:Cancel
*
Shutdown
*
Aspirate CELLCLEAN.
It will take approx. 5 minutes.
Ready
Figure 2-9-1: Shutdown Screen
NOTE: When [3] key is pressed, shutdown is canceled and the Analysis
screen returns.