S
ERVICE
M
ANUAL
S
YSTEC
V/D
S
ERIES
0
51
-X
V
051
-XD
Systec GmbH
Konrad-Adenauer-Straße 15
D-35440 Linden, Germany
Tel.: +49(0)6403 67070-0
Fax.: +49(0)6403 67070-222
01.01.2015
Revision: 2.1

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Systec DE-45 and is the answer not in the manual?

Summary

1 NOTES ON THE DOCUMENTATION

1.2 Safety instructions

Details essential safety guidelines for operating the device and highlights potential hazards.

2 COMMISSIONING

2.9 Setting parameters

Explains how to set various parameters using access levels and codes.

3 MECHANICAL ASSEMBLIES

3.3 Door lock

Explains the automatic door locking mechanism, its principle, and operation.

4 ELECTRICAL ASSEMBLIES

5 REGULAR TESTS AND MAINTENANCE

5.2 Calibration and adjustment

Explains how to calibrate analogue inputs using service software or the control unit.

6 SERVICE SOFTWARE

6.2.2 Setting the communication parameters

Guides users on configuring communication parameters for connecting to the device.

6.2.5 Download Main

Explains how to update firmware on the digital circuit board using S19 files.

6.2.10 Parameters

Details how to view and modify device parameters, including cycle and global settings.

7 TROUBLESHOOTING

7.1 Reset

Provides instructions and safety precautions for resetting the controller software.

7.3 Troubleshooting

Offers a guide with troubleshooting diagrams to locate and resolve errors step-by-step.

7.3.3 Error message: High Gen.Press

Troubleshoots high pressure errors in the steam generator, checking sensors and parameters.

7.3.4 Error message: Water pump error

Diagnoses water pump errors, focusing on electrode defects and steam generator checks.

7.3.5 Error message: No chamb water

Troubleshoots 'No chamb water' errors by checking the water inlet valve and electrodes.

7.3.6 Error message: No Demin.Water

Addresses 'No Demin.Water' errors by checking water level electrodes and steam generator inlet.

7.3.7 Error message: Low steam

Troubleshoots low steam errors by checking steam generator functionality and pressure parameters.

7.3.8 Error message: No Gen.Water

Diagnoses 'No Gen.Water' errors related to the low electrode on the generator.

7.3.9 Error message: No Tap Water

Troubleshoots 'No Tap Water' errors by checking cooling water supply and vacuum unit.

7.3.10 Error message: Low vacuum

Addresses low vacuum errors by checking vacuum pump operation, wear, and leaks.

7.3.11 Error message: Comp.air error

Troubleshoots compressed air errors by checking pressure, valves, and air supply.

7.3.12 Error message: Ring sw. error

Diagnoses ring switch errors by checking air supply, valves, scissor mechanism, and connections.

7.3.14 Error message: Sensor Error

Guides on resolving sensor errors by checking/replacing sensors, wires, or analogue boards.

7.3.15 Error message on D-23: Door lock error

Troubleshoots D-23 door lock errors related to the 'Door Locked' digital input.

7.3.16 Error message on D-23: Door unlock error

Troubleshoots D-23 door unlock errors related to the 'Door Unlock' digital input.

7.3.17 Error message on D-23: Door switch error

Diagnoses D-23 door switch errors related to the 'Door Closed' digital input.

8 PARAMETERS

Systec DE-45 Specifications

General IconGeneral
ModelDE-45
ManufacturerSystec
Chamber Volume45 liters
Sterilization ProgramsMultiple customizable programs
Safety Featurestemperature monitoring
Power Supply230V, 50/60Hz
MaterialStainless steel chamber