EC900 Series Operation Manual V4.7
Technical Specification contd.
All gases except corrosive gases
All gases except corrosive and/or
acidic gases
4 Digit high visibility LED
1/8” OD compression (Swagelok)
-5°C to 50°C (Instrument and sample gas)
Sample Gas Inlet
Pressure:
3.5 – 29 PSI (0.25 Bar - 2 Bar)
Ambient (for pump equipped systems)
Refer to the drawing B900 176 at the rear of the manual Appendix 1.
OPTIONS
Two with volt free c/o contacts fully
configurable over the complete oxygen range,
one fault alarm.
Contact rating 5A @ 250VAC/30VDC
Software configurable over the entire oxygen range for either :
Current: 4-20mA isolated
Current 0-20mA isolated
Auto Calibration
(option):
Allows connection to a cylinder of gas with a certified level of oxygen.
Analyzer will correct to recorded level at programmed intervals.
Improves the response time of the analyzer due to sample transport lag by
adding a fast side stream.
Allows the instrument to be used in situations where there is no pressure to
the sample gas.
Allows the instrument to be used in situations where the sample gas will
react with either brass or copper fittings and pipes. Note: stainless steel
flowmeter and regulator will also be required.