EC900 Series Operation Manual V4.7
13.2 Calibration with Certified Gas
Connect the certified gas to the sample port ensuring the same pressure as the sample gas,
in order to obtain the same flow rate as the sample gas, allow the reading to stabilise.
Enter the SET-UP mode by pressing
Change the CAL value from the default to the value printed on the certification label on the
cylinder, making sure you have changed the %/ppm LED, if appropriate
Pressing the ENTER button
will force the display to the value set
for the certified bottle.
To return to oxygen measurement.
See Front Panel Guide.
Note:- If the CAL value is correct press
to calibrate at
that level
Connect the certified gas to the sample port ensuring the same pressure as the sample gas,
in order to obtain the same flow rate as the sample gas, allow the reading to stabilise
Enter the SET-UP mode by pressing
the programme button
on the front panel of the instrument
If the certified gas is above 2,000ppm enter the value in SPHi. If it is below 2,000ppm enter
the value in SPLo, making sure you have changed the %/ppm LED, if appropriate
Pressing the ENTER button
will force the display to the value set
for the certified bottle
button to return to oxygen
measurement.
See Front Panel Guide
Note:- If the CAL value is correct press
to calibrate at
that level.