Preparations for Operation
Immobilizing the Stand / Arm
To prevent shaking during operation, immobilize the stand and arm prior to operation.
To immobilize the stand and arm, turn the various locking handles in the clockwise direction until they stop.
1.Immobolize the base by lowering the stoppers
on the base's two stopper-attached casters.
OM8-T26
LowertoLock
OM8-T51
Counter-balancedarmswivellockinghandle
Verticalmovementlockinghandle
Lamparmswivel
lockinghandle
2.If necessary, turn the counter-balance arm's
and the lamp arm swivel locking handles 
clockwise until they stop, locking the arms.
Centering
Prior to operation, set the microscope unit at the center of its vertical, lateral and longitudinal strokes (its
range of movement) according to the following instructions.
・In case of using Type Ⅰ Foot switch unit.
1. Step on the front or the back of the foot switch
unit's focus switch to rise or desend the
microscopeso that the center mark on the
microscope comes to the center of the stroke
mark.
OM8-T52
Focusswitch
fordescent
Focusswitch
forrise
OFF
RED
REFLEX
Strokemark