34
Replacing Consumables
Notes on Replacing Consumables
● Do not replace consumables unless you are experienced with the instrument.
● Turn the instrument power o󱐯 , and remove the power cord from the wall outlet prior to proceeding.
● Do not pull the cord itself when removing the power cord.
● Be sure that your hands are not wet.
● Use only appropriate tools.
● Replace the consumables correctly by following the procedure below.
Replacing the Lamp
Be sure to maintain the lamps in both lamp units working. That way one lamp can be used as a
backup in case the other lamp burns out during operation.
● Before replacing the lamp, turn o󱐯 the power and wait for the lamp to cool.
* The instrument and the lamps are hot immediately after use; burn injuries may result.
● Use lamps with the ratings (O91-02 : 15V, 150W) designated by our company. Do not use
any others.
Burn injuries or breakage of the OM-8 may result.
1.Turn o󱐯 the power switch and pull the power
cable plug out of the socket.
・If the lamp was just in use, turn off the
microscope light intensity control knob and
let the lamp cool for 15 minutes.
OM8-T67
Takeout
OM8-T68
Takeout
2.Take out the lamp unit from the counter-
balanced arm.
3.Remove the bulb retaining spring from the
hook and take out the lamp together with
the lamp socket.
*If the lamp retaining spring is stretched too
much, it will lose its resilience.
CAUTION