36
Maintenance and Servicing / Storage and Transport
■ Caring for the Instrument
● Turn the instrument power o , and remove the power cord from the wall outlet prior to
proceeding.
● Do not pull the cord itself when removing the power cord.
● Be careful not to get water on the power cable plug, the fuse holder area, the arm area, or the
microscope unit. It could cause maifunctioning or some other problems.
・Remove the dirt from the main unit's surface by wiping it o with a sort, dry, clean cloth that will not leave
any bers.
*Whenever a "cloth" is mentioned in the procedures below, it means a cloth with the qualities mentioned
above.
・If there is dirt that can't be wiped o with a dry cloth, wipe it o with a cloth moistened with cold or
lukewarm water, then wipe again with a dry cloth.
・If there is dirt that still won't come o, wipe it o with a cloth moistened with a neutral cleaning agent
thinned with cold or lukewarm water, wipe o the cleaning agent with a cloth moistened with cold or
lukewarm water, then wipe again with a dry cloth.
* Do not use organic solvent (such as thinner, benzine, acetone, toluene, and ethyl acetate) or bleach to
clean the instrument. This may damage the painted surfaces or the resin components.
■ Scheduled Servicing
・This instrument (Operating Microscope) does not have a speci c timetable for scheduled servicing; the
instrument supervisor shall always check the instrument prior to and after use.
■ Storage and Transport
● If storing for an extended period of time, turn the instrument power o , and remove the power
cord from the dedicated wall outlet.
● Do not pull the cord itself when removing the power cord;
* Fire may result.
● Remove any twisting and tension from the cords.
・The OM-8 dislike dust. Therefore, always keep it covered with the dust cover whenever it is not being used.
・Use the packing box provided and store under the following conditions:
・Temperature: between -20 and +60 degrees celsius
・ Humidity: between 10 and 95%
・When intending to move the OM-8, rst tighten all the lock handles and secure all the parts. Then move it.
Strong jolts can aect the functioning of the OM-8 and cause other problems as well
■ Disinfection
● Disinfect the Sterilizable caps before surgery.
・The Sterilizable caps can be disinfected by EOG, or autoclave.
Remove
PowerCable
PowerPlug
CAUTION
CAUTION