47
EMC Declaration of Conformity
Recommended separation distances between
portable and mobile RF communications equipment and the OM-8
Table6
TheOM-8isintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbancesare
controlled.ThecustomerortheuseroftheOM-8canhelppreventelectromagneticinterferencebymaintaininga
minimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andtheOM-8as
recommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.
Ratedmaximumoutputpowerof
transmitter
W
Separationdistanceaccordingtofrequencyoftransmitter
m
150KHzto80MHz
d=1.2 √
P
80MHzto800MHz
d=1.2 √
P
800MHzto2.5GHz
d=2.3 √
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
Fortransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistanced
inmeters(m)canbeestimatedusingtheequationapplicabletothefrequencyofthetransmitter,wherePisthe
maximumoutputpowerratingofthetransmitterinwatts(w)accordingtothetransmittermanufacturer.
Note1: At80MHzand800MHz,theseparationdistanceforthehigherfrequencyrangeapplies.
Note2: Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaectedbyabsorption
andreectionfromstructures,objectsandpeople.