21
Connections,
Linear Control Operation (Analog Out Control Mode)
JP7 JP21 2 3 4 10 9 8 7 6 5 4 3 2 1
(+)
(-)
TE DEVICE
TE DEVICE
PROGRAMMABLE POWER
SUPPLY
OPTIONAL FUSE
THERMOSTAT AND OTHER
PROTECTIVE DEVICES
(CUSTOMER SUPPLIED)
OPTIONAL
SECONDARY
SENSOR
JP2-4 (+) and
JP2-3 (-)
CONTROL
SENSOR
JP2-1 (+) and
JP2-2 (-)
12 V, 36 V, 20 A maximum
JP7
PIN 1
JP2
PIN 1
WIRES MUST PASS THROUGH
GASKET HOLES
NOTE: LID IS REMOVED TO SHOW
WIRE CONNECTIONS
USB COMMUNICATION PORT
CONTROLLER
POWER SUPPLY
12 V, 36 V, 150 mA minimum
(-)
(+)
to JP2-7
to JP2-9
to JP7-1
to JP7-2
to JP7-3
to JP7-4
remote control (+)
remote control (-)
to JP2-5