CSBCryoEMFacilityTF20OperatingInstructionsSEC
05/02/16
IV.EndofSession
1. Closecolumnvalves
2. Retractthecampera
3. Putthescreendown
4. Home the stage (go to SearchStage and select the tab Navigate to Control and click on
Holder,thisensuresthatthestageisat0uminX,YandZ.
5. Removetheholdercarefully.
6. Clickoncryocycle.Makesurethecryocyclewillrunfor365minutesormore.
Wait~5minutesorforTurbototurnyellowbeforeremovingtheliquidnitrogendewar
7. Dismountthespecimenfromtheholderanddisposeofit(orsaveit)properly.
8. Once the Turbo turns yellow, remove the dewar from the microscope and dispose of the
remainingLN2inthefoamcontainerunderthetable.
9. Placetheprotectivefabricunderneaththeanticontaminator.
10. IfusingtheRTorRTtomographyholders,returnthemtotheirproperreceptacle.
11. If using the Gatan cryo holder(s), return to the dry pumping station and follow the instructions
thereonhowtowarmitandpumpitproperly.
12. Pleasesignupontheuser'ssheet.Recordtheactualtimespentimaging.
13. Return the next day to retrieve Data. Data will be moved overnight to the computer located
behindthecurtainintheF20room
V. Imaging
1. OpenDigitalMicrograph
2. Setexposuretimeto1second
3. Wantthemeanvaluetoequal~1000