16 USING SAMPLE FILTER FOR CHECKS
This operation must be performed monthly or when one thinks that the opacimeter module is not
performing correct measurements.
Open the back of the instrument.
Insert the sample filter and check that the
instrument reads the value on the filter with a
tolerance of ±1%.
Close the back of the instrument.