System Installation
IR310 Staff Station: The IR310 provides a communication point for staff members to call the master sta-
tion for staff to staff communication. The device also has a special feature that is activated by turning on
dip switch 7. This feature allows the device to call in as a patient station for applications in which call
cords are not desired for patient call in. A call button, reset button, call-placed indicator and in-use
status indicator are provided. See IR3xx-series Station—General Information on page 9 for additional
details.
Call Cords:
SF301/SF302-series Single and Dual push button call cord, 7′ or 10' available (for SF380)
SF311 Push button call cord for 8P8C jack, 7' or 10' available (for IR311, IR312)
SF401 Push button call cord, 7′ (for IR319-series, IR320-series)
SF41-series Pillow speaker, 8′ cord, intercom & controls for lights and TV (for IR311-series
with PM311C, PM321; IR312-series with PM312C, PM322B)
SF401KIRL Pillow speaker, 8' DIN plug, 2 light switches, intercom, channel keypad and IR
NOTE: Other options for geriatric call cords are available. Contact Sales for details.
Call cords are inserted into their associated patient stations and are subject to the same environmental
requirements.
LI38x/LI38xLED Dome Light: The LI380 provides visual indication from the IR3xx-series stations and
associated peripheral devices. Four bulbs and four colored lenses (LI380), four colored LEDs and single
lens (LI380LED) allow the device to indicate all system call types with various flash rates and com-
binations. When connected to an IR315-series Duty Station, the light functions as a zone light. The LI
series dome lights use the Steel City H2BD backbox with a 2GC plaster ring; refer to Figure 2 on page
8 for mounting. The minimum dimensions for the backbox must be not less than 4"×4"×1½", and the
minimum clearance from current carrying parts to dead metal parts must be no less than 0.5". The dome
light must be located above or beside the doorway of the associated room, and must be mounted so that
unobstructed visibility is provided. When being used as a zone light, the LI380 must be mounted at cor-
ridor heads or junctions so that facility staff members can readily identify the associated zone area. The
LI380 must be within 50' of the patient station it is connected to. The LI380 requires a specific jumper
connection to be made (from pin 1 of the 8-pin connector to pin 6 of the 15-pin connector on the asso-
ciated patient station) to disable the lamp fault detection circuitry of the associated IR3xx-series station
it is connected to (refer to Figure 21 on page 36). The operating environment for the LI380 is 10-40°C
with relative humidity not exceeding 80%. See Wiring Installation on page 12 for additional details.
LI386 Dome/Zone Light: The LI386 provides the same indication functionality as the LI380 and
includes support electronics to make it an addressable device. This device is suitable for stand-alone
equipment applications and zone indication applications. NOTE: This device does not support all of the
standard peripheral devices (refer to Figure 23 on page 38). Turning on dip switch 7 activates the zone
indication feature. See LI380 Dome Light (above) for location and environmental requirements.
SF-series Peripheral Devices—General Information: The SF-series peripheral devices’ electronics are
mounted on flame retardant ABS plastic faceplates. SF-series devices must be mounted within 50' of the
patient station to which it will be connected. They are mounted on single-gang rings or boxes (minimum
ring opening not to be less than 1¾" × 2¾"). The minimum clearance from current carrying parts to dead
metal parts must be no less than 0.5". Plug-on connectors are provided for easy installation. See the Wir-
ing Installation on page 12 for additional details.