CALIBRATION
PROCEDURE
067-0587-01
the
individual
gain
ranges,
these
voltages
can
be
measured
accurately.
These
precision
dividers
are,
however, entirely
dependent
upon
the
+50
volt
and
-50
volt
power
supplies
generated
by
the
oscilloscope
mainframe.
Errors
in
these
power
supplies
must,
therefore,
be
taken
into
account
be-
fore
arriving
at
a
final
answer.
The
following
formula
is
used
to
correct
the
reading
obtained
in
Step
9.
For
power
supply
errors:
Range
X
(E+ps
+
E-ps)
Corrected
Reading
=
Reading-
200
Where:
Range
is
the
staircase
step
being
checked
(50
mV,
100
mV......)
Etps
is
the
error
of
the
+
50
volt
power
supply
stated
in
per
cent,
E-ps
is
the
error
of
the
-
50
volts
power
supply
stated
in
per
cent.
Example:
A
reading
of
+100.35
mV
is
obtained
on
the
100
mV
range.
A
+0.4%
error
is
noted
for
the
+50
volt
supply
and
a
-0.1%
error
is
noted
for
the
-50
volt
supply.
The
corrected
reading
would
be:
100.35
my
-
200
av
o.4
—0-1)
.
100.35
mV
-
0.15
mV
=100.20
mV
Procedure
a.
Plug
the
signal
pickoff
device
into
the
Signal
Standardizer.
Place
the
Signal
Pickoff/Signal
Standardizer
on
the
Flexible
Extender,
b.
Set
the
Signal
Standardizer
VERT
or
HORIZ
switch
to
GAIN.
c.
Measure
and
calculate
the
%
error
of
the
+50
volt
power
supply
and
~-50
volt
power
supply.
Record
these
errors
for
future
use.
A
conven-
ient
point
to
place
the
DVM
probe
for
measuring
the
+50
volt
power
supply
ig
ona
pin
4-19
of
the
plug-in
rear
connector.
The
-50
volt:supply
is
located
at
the
pin
B-19.
4-7