CALIBRATION
PROCEDURE
067-0587-01
4-6
PROCEDURE
i.
REP
RATE
-—
Check
a.
Check
that
the
frequency
of
the
displayed
waveform
is
1
mega-
hertz
+200
kilogertz
(period
of
1uS
+0.2y8).
b.
Check
that
the
frequency
of
all
REP
RATE
switch
positions
are
within
the
tolerances
listed
below.
TABLE
1
REP
RATE
TOLERANCE
1
MHz
+200
kHz
(1uS
+2uS)
100
kHz
+20
kHz
(10uS
+2yuS)
10
kHz
+2
kHz
(100uS
+20yS)
1
kHz
+200
Hz
(lms
+200,uS)
100
hz
+20
Hz
(10ms
+2ms)
c.
Return
REP
RATE
switch
to
1
MHz.
2.
Amplitude
DC
Balance
a.
Place
the
Signal
Standardizer
on
the
Flexible
Extender
‘and
remove
the
plug-in
right
side
cover.
b.
Set
the
Signal
Standardizer
VERT
or
HORIZ
switch
to
AUX
IN.
c.
Adjust
R25
for
minimum
trace
shift
while
rotating
the
AMPLITUDE
control
throughout
its
range.
3.
Gain
Accuracy
(Check)
Theory:
The
067-0587-01
gain
circuitry
uses
precision
dividers
to
develop
the
desired
voltages
across
the
oscilloscope
mainframe
50
{
res-
istor.
By
substituting
precision
502
resistors
and
by
stepping
through

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 067-0587-01 and is the answer not in the manual?

Tektronix 067-0587-01 Specifications

General IconGeneral
BrandTektronix
Model067-0587-01
CategoryTest Equipment
LanguageEnglish

Related product manuals