CALIBRATION
PROCEDURE
067-0587-01
g.
Set the
Signal
Standardizer
AMPLITUDE
control
to
obtain
a
4-
division
display.
h.
Adjust
R31
to
just
turn
CW
LEVELED
light
on.
If CW
LEVELED
light
doesn't
go
off,
mechanically
center
R31
and
check
that
the
light
remains
on
while
rotating
the
AMPLITUDE
control
to
obtain
from
four
(4)
to
eight
(8)
divisions.
CW
Leveled
Check
i.
Set the
Signal
Standardizer
AMPLITUDE
control
to
obtain
a
6—division
display.
j.
Increase
the
Constant
Amplitude
Signal
Generator
amplitude
to
1
volt
p-p.
(Check
that
the
display
maintains
approximately
6
divisions
and
that
the
CW
LEVELED
light
remains
on.
k.
Increase
the
Constant
Amplitude
Signal
Generator
to
3
volts
p-p.
Check
that
the
display
is
distorted
and
the
CW
LEVELED
light
extinquishes.
7.
Frequency
Response
Check
a.
Set
the
constant
amplitude
generator
to
0.5
volts
and
adjust
the
Signal
Standardizer
AMPLITUDE
control
to
obtain
a
6—division
display.
b.
Set
the
sampling
sweep
unit
to
2
ns/division.
ec.
Check
that
the
display
amplitude
remains
within
5.82
divisions
to
6.18
divisions
(+3%)
while
varying
the
frequency
of
the
constant
amplitude
signal*generator
from
70
MHz
to
500
MHz.
d.
Remove
all
cables
and
test
equipment.
This
completes
the
calibration
procedure
for
the
067-0587-01
Signal
Stand-
ardizer.
4-15