EasyManua.ls Logo

Tektronix 1910 - Test Signals - Full Field Output (Sync & Burst) Specifications

Tektronix 1910
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table
1·5
TEST
SIGNALS-FULL
FIELD OUTPUT
(Sync
&
Burst)
Specification-1910
Operators
Performance
Supplemental
Perf.
Ck.
Characteristics
Requirements
Information
Step. No.
Amplitude 1
.2
V maximum p-p into 75 Q
Sync
285.7
mV
±2
mV
Peak
Level
714.3mV
±5mV
28
Blanking
Level
DAC
DC
Restorer On oV
±2
mV
40
DAC
DC
Restorer Off oV
±50
mV
40
Field Period 16.68 ms
Digitally determined from 14.3 MHz.
Line Period
63.56
itS
-
Sync Rise &
Fall
Time 140 ns ± 15
ns
10% to 90% amplitude.
Sync Timing
See
Fig.
1-7.
Front Porch
1.7
j.LS
± 100
ns
using a
Digitally determined from 14.3 MHz.
100 IRE pedestal
Line Blanking Interval 11.28
itS
± 100 ns at 20 IRE points
using a100 IRE Pedestal
Breezeway
0.6
j.LS,
±
50
ns, at 50%
of
sync
to
50%
of burst amplitude
Back Porch Duration
4.83
itS,
± 50 ns.
at
50% of sync
to
20
IRE using a 100 IRE pedestal
Line Sync 4.7
itS,
± 50 ns,
at
50%
amplitude
point
Vertical Serration
See
Fig.
1-8.
REV
OCT
1984
1-29