Tektronix TCP312 Service Manual

Tektronix TCP312
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
xx
TCPA300/400 Ampliers &
TCP300/400 Series AC/DC Current Probes
ZZZ
Service Manual
*P077032300*
077-0323-00

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Tektronix TCP312 and is the answer not in the manual?

Tektronix TCP312 Specifications

General IconGeneral
TypeAC/DC Current Probe
Bandwidth100 MHz
Maximum DC Current30 A
Maximum Pulsed Current50 A
Rise Time3.5 ns
Input ConnectorBNC
Sensitivity100 mV/A
ConnectorTektronix Type
Cable Length1.5 m
Power RequirementsPowered by TekConnect® interface
Operating Temperature0°C to +50°C
Maximum Voltage600 V

Summary

General Safety Summary

To Avoid Fire or Personal Injury

Safety precautions to prevent fire or personal injury during operation and handling.

Observe All Terminal Ratings

Adhere to all product ratings and markings to prevent hazards.

Avoid Exposed Circuitry

Do not touch exposed connections and components when power is present.

Service Safety Summary

Do Not Service Alone

Perform service only when another qualified person is present.

Disconnect Power

Switch off power and disconnect the cord before servicing.

TCPA300 and TCPA400 Performance Verification

TCP305 and TCP312 Performance Verification

TCP303 and TCP404XL Performance Verification

TCPA300 and TCPA400 Amplifier Adjustments

TCP305 and TCP312 Adjustments

TCP305 and TCP312 DC Gain Adjustment

Procedure to adjust DC gain for TCP305/312 probes.

TCP303 and TCP404XL Adjustments

TCP303 and TCP404XL Transient Response and Coarse Gain Adjustment

Adjusts transient response and coarse gain for TCP303/404XL.