TCPA300 and TCPA400 Performance Verification
TCPA300 Ampli
fier Test Record
Photocopy this form and use it to record the performance test results.
TCPA300 test record
Instrument Serial Number: Certificate Number:
Temperature
:
Relative Humidity %:
Date of Calibration:
Technician
:
Performance test Range, A/V Minimum Measured/calculated Maximum
Front-Pan
el Display
Pass/Fail
1
-1% +1%
5
-1% +1%
10
-1% +1%
50
-1% +1%
DC Gain Accuracy
(% Error)
50 (CO MP ) -1% +1%
1 .707
——
1 (100
MHz)
.707
——
5
.707
——
10 .707
——
50 .707
——
Bandw
idth
50 (
COMP)
.707
——
TEK
PROBE Cable
1 1.05 Ap-p 1.5 Ap-p
AC C
oupling
1
BN
C Coax Cable
1 1.05 Vp-p 1.5 Vp-p
Degauss
Pa
ss/Fail
C
urrent Overload
P
ass/Fail
1
The AC Coupling test can be performed with either cable, but the units of measure for the results differ between the two.
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Service Manual 17