TCPA300 and TCPA400 Performance Verication
TCPA300 and TC
PA400 Performance Verication
This section
covers the performance verication for the TCPA300 and TCPA400
Current Probe Ampliers. All of the procedures cover both ampliers, exce pt the
Current Overload test, which is for the TCPA300 only. Power o n the amplier and
test equipment and let them warm up for 20 minutes.
Before starting these procedures, photocopy the appropriate test record to record
the performance test results. The recommended calibration interval is one year.
(See page 17, TCPA300 Amplier Test Record.)
Equipmen
tRequired
The performance verication procedures require the listed test equipment. The
test equipment must meet or exceed the specications listed. You m ay need to
modify t
he test procedures if you do not use the recommended equipment.
Table 3: Required test equipment
Qty Item Description Recommended
1
Oscilloscope
500 MHz bandwidth
Tektronix TDS5000
1
Leveled Sine Wave Generator
3 MHz to 100 MHz
Wavetek 9100 with >250 MHz Oscilloscope
Calibration Option
1 Digital Multimeter
DCV: 0.1% Accuracy, digit
resolution
Keithley 2700
1
Current Source DCA: 0.1% Accuracy, 0 to
±100 mA, oating outputs
Fluke 5100 or HP 6612C
1
Current Source DCA: 0.1% accuracy, 0 to ±15 A
Wavetek 9100
1
Square Wave Generator ACV 1.5 Vp-p, 28 Hz square wave
Wavetek 9100
1
HF Current Loop 50 , BNC Connector
Tektronix part number 015-0601-50
1 Precision Termination
50 ±0.1%, 0.5 W
Tektronix part number 011-0129-00
2
BNC Cables 50 , 0.76 m (30 in) long
Tektronix part number 012-0117-00
1
Cable TEKPROBE Interface Cable
Tektronix part number 012-1605-00
2 Adapter
BNC-to-Dual Banana
Tektronix part number 103-0090-00
1
DC Current Loop 5 turns 18 AWG coated wire (See page 4, Making DC Current Loops.)
1
DC Current Loop 50 turns 13 AWG coated wire (See page 4, Making DC Current Loops .)
1
Calibration Adapter TCPA Calibration Adapter
Tektronix part number 174-4765-00
1
Current Probe
1
TCPA300/400-compatible probe Tektronix TCP303, 305, or 312 (TCPA300)
or TCP404X L (TCPA400)
1
Required for functional test only.
TCPA300/400 Ampliers and TCP300/400 Series Current Probes Service Manual 3