EasyManua.ls Logo

Tektronix TCP312 - DC Gain Accuracy; Figure 2: Equipment Setup for DC Gain Accuracy Test

Tektronix TCP312
120 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TCPA300 and TCPA400 Performance Verication
DC Gain Accura
cy
Use the following procedure to verify the DC gain accuracy of the amplier.
Test Equipment Setup
Set up the test equipment as follows:
1. Connect the output of the amplier to the DMM:
a. Connect a BN
C cable from the amplier OUTPUT to the 50 precision
feedthrough termination.
b. Attach the
termination to a BNC-to-dual banana adapter.
c. Observe polarity and insert the dual banana adapter into the digital
multimet
er DC voltage input.
2. Connect the probe input to the current source:
a. Connect the BNC end of the Calibration Adapter to a BNC-to-dual banana
adapter.
b. Observe polarity and insert the dual banana adapter into the current source
DC output. Do not connect the Calibration Adapter to the amplier at
this t
ime.
Figure 2: Equipment setup for DC gain accuracy test
6 TCPA300/400 Ampliers and TCP300/400 Series Current Probes Service Manual

Table of Contents

Related product manuals