TCP303 and TCP404XL Performance Verification
TCP404XL Curr
ent Probe Test Record
Photocopy this form and use it to record the performance test results.
TCP404XL test record
Instrument Serial Number: Certificate Number:
Temperature
:
Relative Humidity %:
Date of Calibration:
Technician
:
Performance test Range Test Minimum Measured/calculated Maximum
3.0 A
-3% +3%
9.0 A
-3% +3%
DC Gain Accuracy (% Error) 1 A/mV
15.0 A
-3% +3%
Rise Tim
e
1A/mV
——
175 ns
Bandwi
dth
1A/mV
—
2MHz
——
TCPA300/400 Amplifiers and TCP300/400 Series Current Probes Service Manual 43