Operating basics
Maximum leadin
gedgetime. This value is the minimum of the three in each
instance.
If runMode = Co
ntinuous:
Temp1=0.8*2.0*width–tEdge;
Temp2=(period–width)*0.8*2.0–tEdge;
Temp3 = 0.625 * period.
Else:
Temp1=0.8*2.0*width–tEdge;
Temp2 = ( p
eriod leadDelay width ) * 0.8 * 2.0 tEdge;
Temp3 = 0.625 * period.
Maximum trailing edge time. This value is the minimum of the three in each
instance.
If runMode = Continuous:
Temp1
=0.8*2.0*width–lEdge;
Temp2=(period–width)*0.8*2.0–lEdge;
Temp3 = 0.625 * period.
Else:
Temp1=0.8*2.0*width–lEdge;
Temp2=(period–leadDelay–width)*0.8*2.0–lEdge;
Tem
p3 = 0.625 * p eriod.