Performance Verification
AFG3000 Series Arbitrary/Function Generators Service Manual 4-13
AFG325x Spectrum Analyzer reading
Sine 1 MHz 1 MHz 2 MHz 3 MHz 4 MHz 5 MHz
CH1 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –50 dBc
CH2 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –50 dBc
Sine 5 MHz 5 MHz 10 MHz 15 MHz 20 MHz 25 MHz
CH1 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –37 dBc
CH2 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –37 dBc
Sine 100 MHz 100 MHz 200 MHz 300 MHz 400 MHz 500 MHz
CH1 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –30 dBc
CH2 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –30 dBc
Sine 240 MHz 240 MHz 480 MHz 720 MHz 960 MHz 1.20 GHz
CH1 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –30 dBc
CH2 Harmonic Distortion dBc dBc dBc dBc dBc
reading – reference 0 dBc dBc dBc dBc dBc Nth – reference < –30 dBc
AFG3000 Series Performance Test Record
Harmonic Distortion Fundamental
= reference
2nd 3rd 4th 5th Limit