Brief procedures
8. Click the Reset
to default setup button in the toolbar.
9. Set the function generator output to square wave, 1 kHz, 5 V
p-p
.
10. Turn on the output of the function generator.
11. On the instrument, load the PV_Square.wfmx waveform as an output
waveform. Follow the steps below:
a. In the Waveform List window, click Open Waveform and navigate to
C:\Program Files\Tektronix\AWG70000\Samples\PV.
b. Select Open File.
c. In the Waveform List window, select (drag and drop) the
PV_Square.wfmx waveform on the work space area.
12. In the Run tab on the bottom of the screen, sele ct Triggered, and set the
Trigger Input to A.
13. Click the CH1 Enable button.
14. In the Setup tab, click Trigger, and set the trigger level to 1.0 V, Rising, 50 Ω.
15. Click
the Play button on-screen or on the instrument.
16. Press the All Outputs button on the instrument to output the waveform.
17. Verify that the output is displayed on the AWG work space.
18. Verify that the output is displayed on the oscilloscope.
19. Repeat the test for the Trigger B input: Move the cable from the Trigger A
input to the Trigger B input, and then under the Run tab, set the trigger input
to B.
20. Verify that the Trigger B output is displayed on the AWG work space.
21. Verify that the output is displayed on the oscilloscope.
22. Disconnect the test se tup.