EasyManua.ls Logo

Tektronix CT-1

Tektronix CT-1
27 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
When droop is pr
esent at low frequencies, the apparent overall peak-to-peak
height from top to bottom is not the true current. The height of a at-top pulse still
can be measured accurately by observing the transition edge of the pulse. The
50 mA pulse is faithfully reproduced at the high-to-low transition at the center
of the screen. (See Figure 7.)
Pulse width, tilt, and the lower 3 dB frequency are related by the formula:
Percent tilt = 200π Tp f
1
Where:
Tp = pulse w
idth
f
1
= l ower 3 dB frequency
Figure 7: CT-2 decay characteristics referenced to transition
Additional consideration needs to be given to the amp-second rating of the
tran
sformer. The product of the pulse current and the pulse duration should be less
than the amp-sec ond ra ting to kee p the transformer core from saturating. When
saturation occurs, the output voltage falls to zero.
Probe Handling
Only normal handling considerations are necessary with the CT-1 and CT-2
transformers. They are sealed units and are not designed to be disassembled. If
the transformer or cable is damaged, return it to Tektronix for replacement.
CAUTION. Dropping the CT-1 or CT-2 transformer may break the probe. Do not
p
ull or stretch the P6041 cable or place objects on the cable. This may damage
the probe cable.
CT-1 and CT-2 Current Transformer Instructions 7

Other manuals for Tektronix CT-1