Commands Listed in Alphabetical Order
WFMInpre:BN_Fmt
Sets or returns the format of binary data for incoming waveforms.
Group
Waveform Transfer
Syntax
WFMInpre:BN_Fmt {RI|RP}
WFMInpre:BN_Fmt?
Related Commands
WFMOutpre:BN_Fmt
Arguments
RI species signed integer data point representation.
RP species positive integer data point representation.
Examples
WFMINPRE:BN_FMT RP species positive integer data point representation.
WFMINPRE:BN_FMT? might return :WFMINPRE:BN_FMT RI indicating that the
incoming data is currently interpreted as signed integers.
WFMInpre:BYT_Nr
Sets or returns the data width for the incoming waveform. Changing the value of
WFMInpre:BYT_Nr also changes the value of WFMInpre:BIT_Nr.
Group
Waveform Transfer
Syntax
WFMInpre:BYT_Nr <NR1>
WFMInpre:BYT_Nr?
Related Commands
WFMInpre:BIT_Nr
Arguments
<NR1> is the number of bytes per data point and can be 1 or 2.
Examples
WFMINPRE:BYT_NR 1 sets the number of bytes per incoming waveform data
point to 1, which is the default setting.
WFMINPRE:BYT_NR? might return :WFMINPRE:BYT_NR 2 indicating that there
are 2 bytes per incoming waveform data point.
2-354 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012B and is the answer not in the manual?

Tektronix DPO2012B Specifications

General IconGeneral
BrandTektronix
ModelDPO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals