g. Check - The measurement is within the limits specified in the test record.
3. Set the Vertical Scale to 100 mV (140 mV for >20 GHz models) per division
and repeat step 2.
4. Repeat steps 2through 3for the remaining input channels:
a. Move the test setup to an unchecked input channel.
b. Set the Vertical Scale of the channel to 10 mV (6.25 mv for >20 GHz
models) per division.
c. Repeat steps 2 through 3.
5. Disconnect the hookup: Disconnect the equipment from the instrument.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)