Analyzing Wavef
orms
6. You can view tim
e-domain and
frequency-domain waveforms
simultaneously.
Use Gating to s
elect only a portion of
the time-domain waveform for spectral
analysis. (See page 85, Gating.)
Quick Tips
Sources for spectral math waveforms must be channel or other math waveforms.
Use short record lengths for faster instrument response.
Use long record lengths to lower the noise relative to the signal and increase the frequency resolution.
Different window functions produce d ifferent lter response shapes in the spectrum and result in different resolution
bandwidths. See the online help for more information.
The resolution bandwidth (RBW) directly controls the gate width. Therefore, the time domain gate markers move
as you adjust the control.
You can display the linear magnitude of the real data or the imaginary data in the spectrum. This is useful if you process
the spectrum off line and transform it back into a time domain trace.
Using Mask Testing
Serial Communications Mask Testing (Option MTM or MTH) allows you to compare your signal to a predened template
or mask. For the signal to pass the test, it must fall outside the segments dened by the mask. Generally, standards
commi
ttees such as ANSI dene the masks. To perform mask testing do the following:
1. Select Masks > Source....
98 DPO7000 Series and DPO /DS A70000/B Series Quick Start User Manual

Table of Contents

Other manuals for Tektronix DPO7354

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO7354 and is the answer not in the manual?

Tektronix DPO7354 Specifications

General IconGeneral
BrandTektronix
ModelDPO7354
CategoryTest Equipment
LanguageEnglish

Related product manuals