Integration Algorithm. The integration algorithm used by the instrument is as follows:
W(t) is the sampled waveform
Å´(t) is the continuous function obtained by linear interpolation of W(t)
A and B are numbers between 0.0 and RecordLength – 1.0
If A and B are integers, then:
where s is the sample interval.
Similarly,
W(t) is the sampled waveform
Å´(t) is the continuous function obtained by linear interpolation of W(t)
A and B are numbers between 0.0 and RecordLength – 1.0
If A and B are integers, then:
where s is the sample interval.