EasyManuals Logo

Tektronix MDO4104C User Manual

Tektronix MDO4104C
21 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
Built-In Security Features
To Overwrite Protected
User Data
This procedure
overwrites everything that is stored in the protected user data area.
Send the following comm ands to the oscilloscope:
:PASSWORD “XYZZY” (or current password if changed from the default of
“XYZZY”)
:NEWPASS “ANYTHING”
:PASSWORD “ANYTHING”
*PUD #3 300 xxx xxx xxx xxx xxx xxx xxx xxxx xxx xxx xxxxxxxxxx[repe at
for a total of 300 x’s]
NOTE. If you do not have access to a program that supports sending programmatic
commands to the instrument, copy the preceding commands to a text le that ends
in “.set
and recall the le from the Recall Setup menu.
For more information on using programming commands, refer to the Programmer
Manual
for this instrument, available at www.tektronix.com/downloads.
10 Declassication and Security Instructions

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4104C and is the answer not in the manual?

Tektronix MDO4104C Specifications

General IconGeneral
BrandTektronix
ModelMDO4104C
CategoryTest Equipment
LanguageEnglish

Related product manuals