EasyManuals Logo

Tektronix MDO4104C User Manual

Tektronix MDO4104C
21 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #21 background imageLoading...
Page #21 background image
Clear and Sanitize a Nonfunctional Instrument
If your instrument is not functioning and you need to clear or sanitize it, proceed
as follows:
NOTE. The An
alog board must be calibrated with the Main board by Tektronix.
Please contact Tektronix before returning your instrument.
1. Remove the
USB ash drive from your oscilloscope. Refer to your company's
internal policies regarding handling or disposal of the ash drive.
2. Remove th
e Main board from your oscilloscope. For removal instructions,
refer to the Service manual for the instrument, which is available on the
Tektronix Web site at www.tektronix.com/manuals. Refer to your company's
internal policies regarding handling or disposal of the board.
3. Return the oscilloscope to Tektronix. A new M ain b oard will be installed,
and the oscilloscope will be repaired. The oscilloscope will be adjusted as
necessary, which includes adjusting (calibrating) the new Main board and
the Analog board together.
4. Replacement of any missing hardwa re will be charged according to the rate
at the time of replacement.
Declassication and Security Instructions 13

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4104C and is the answer not in the manual?

Tektronix MDO4104C Specifications

General IconGeneral
BrandTektronix
ModelMDO4104C
CategoryTest Equipment
LanguageEnglish

Related product manuals