Trigger Setup
1. Adjust the position (delay) time by
rotating the Horizontal Position knob.
2. Turn the horizontal Scale knob to acquire
the detail th
at you need around the
position (delay) expansion point.
The part of the record that occurs before the trigger is the pretrigger portion. The part that occurs after the trigger is the
posttrigger portion. Pretrigger data can help you troubleshoot. For example, to find the cause of an unwanted glitch in
your test circuit, you can trigger on the glitch and make the pretrigger period large enough to capture data before the
glitch. By analyzing what happens before the glitch, you may uncover information that helps you find the source of the
glitch. Alternatively, to see what is happening in your system as a result of the trigger event, make the posttrigger period
large enough to capture data after the trigger.
Slope and Level
The slope control determines whether the
oscilloscope finds the trigger point on the
rising or the falling edge of a signal.
The level control determines where on that
edge the trigger point occurs.
The oscilloscope provides a long horizontal
bar or bars across the graticule to temporarily
show the trigger level.
1. Turn the front-panel Trigger Level knob
to ad
just the trigger level without going
to a menu.
2. Push the Force Trig button to forc e the
oscilloscope to trigger.
66 MSO2000B and DPO2000B Series Oscilloscopes User Manual