Performance Verication
12. Set the channel
threshold to +4 V as follows:
a. Push the front-panel D15-D0 button.
b. Push the Thresholds lower-bezel button.
c. Turn the Multipurpose a knob to select channel D0.
d. Push the Fine front-panelbuttontoturnoffthene adjustment.
e. Turn the Multipurpose b knob and set the value near 4.00 V (4 V/div).
f. Push the Fi
ne buttontoturnthene adjus tment on again.
g. Turn the Multipurpose b knob and set the value to exactly 4.00 V
(4 V/div)
.
13. Set the DC voltage source (Vs) to +4.4 V. Wait 3 seconds. Check the logic
level of
the channel D0 signal display.
14. Decrement Vs by -10 mV. Wait 3 seconds and check the logic level of the
channe
l D0 signal display. If it is a static logic low, record the Vs value as
V
s+
in the 4 V row of the test record.
If the
signal level is a logic high or is alternating between high and low, repeat
this step (decrement Vs by 10 mV, wait 3 seconds, and check for a static logic
low) until a value for V
s+
is found.
15. Push the front-panel Trigger Menu button.
16. Clic
kthelower-bezelSlope button to change the slope to Rising.
17. Set the DC voltage source (Vs) to +3.6 V. Wait 3 seconds. Check the logic
lev
el of the channel D0 signal display.
If the signal level is a static logic high, change the DC voltage source Vs to
+3
.5 V.
18. Increment Vs by + 10 mV. Wait 3 seconds and check the logic level of the
c
hannel D0 signal display. If it is a static logic high, record the Vs value as
V
s-
in the 4 V row of the test record.
If the signal level is a logic low or is alternating betwe en high and low, repeat
this step (increment Vs by 10 mV, wait 3 seconds, and check for a static logic
high) until a value for V
s-
is found.
19. Find the average using this formula: V
sAvg
=(V
s-
+V
s+
)/2. Record the average
as the test result in the test record.
MSO4000B and DPO4000B Series Specications and Performance Verication 77

Table of Contents

Other manuals for Tektronix MSO4104B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4104B and is the answer not in the manual?

Tektronix MSO4104B Specifications

General IconGeneral
BrandTektronix
ModelMSO4104B
CategoryTest Equipment
LanguageEnglish

Related product manuals