Performance Verification
Compare the tes
t result to the limits. If the result is between the limits, the
channel passes the test.
20. Repeat the pro
cedure for all remaining digital channels as follows:
a. Push the D15–D0 button.
b. Move the D C voltage source connection, including the ground lead, to the
next digital channel to be tested.
c. Starting from step 3 , repeat the procedure until all 16 digital channels
have been tested.