To prevent electrostatic damage to components whenever you work on the instrument, wear properly-grounded
electrostatic prevention wrist and foot straps, and work in a tested antistatic environment on an antistatic mat.
Remove SA
TA riser board assembly on page 9
Remove rear chassis assembly on page 11
Remove the baffle bracket on page 12
About this task
Steps:
Procedure
1. Use a T-10 Torx screwdriver to remove the screw between the BNC connectors on the rear panel of the chassis. Use a
3/16-inch open end wrench to remove the four jack screws from the video connectors on the rear panel.
2. Disconnect the black cables connecting the power supply to the carrier interface assembly and the white cable on the
AFG riser board. Y
ou do not need to disconnect the AC line filter cables.
Maintenance
6 Series B MSO MSO64B, MSO66B, MSO68B Service 15