i. Repeat substeps 2.c through 2.h until all vertical scale settings, listed in
the table, are checked for the channel under test. (See Table 9: Gain
accuracy on page 213.)
j. Test all channels: Repeat substeps 2.a through 2.i for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the generator output from the channel last tested.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 219