c. Touch the X (close) button.
d. From the button bar, touch Trig, select the A Event tab, and touch the
Select button.
e. Touch the Serial button, set Coding to NRZ, and then set the Standard to
GB Ethernet.
f. Touch the Edit button.
g. Set the Format to Hex and then touch the Clear button.
h. Enter data into the Serial Pattern Data field for one of the settings in the
table that is not yet checked, starting with the first setting. (See Table 19:
Serial pattern data on page 272.)
i. Touch Enter.
j. Right click on the graticule, select Cursors > Move Cursors to Center.
k. Adjust the cursors until the Δt readout equals 800 ps (one unit interval).
l. Center Cursor 2 in the low of the waveform just to the right of the center
graticule line (See Figure 42: Isolated 0 triggering on page 274.)
m. Verify that the instrument triggers one Unit Interval (UI, one baud
divided by the bit period) before the 0 in the input signal. The absolute
value of the T1 cursor readout must be ≤325 ps. Enter pass or fail in the
test record.
n. Touch the Edit and then the Clear button.
o. Enter data into Serial Pattern Data field for the next setting in the table
that is not yet checked. (See Table 19: Serial pattern data on page 272.)
p. Touch Enter.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series 273