3. Record the reference level:
Disconnect the sine wave generator from the instrument.
Connect the sine wave generator to the power sensor.
Note the level meter reading.
4. Set the generator to the new frequency and reference level:
Change the sine wave generator to the desired new frequency.
Input the correction factor and/or the new frequency into the level meter.
Adjust the sine wave generator amplitude until the level meter again
reads the value noted in step 3. The signal amplitude is now correctly set
for the new frequency.
Disconnect the sine wave generator from the power sensor.
Connect the sine wave generator to the instrument.
Performance verification (MSO/DPO70000C, MSO/DPO70000DX, and DPO7000C series)
284 MSO70000C/DX, DPO70000C/DX, DPO7000C, MSO5000/B, DPO5000/B Series