3. Set the Horizontal Delay Mode button to On.
4. Select the number of B events with the Trig Event control.
5. Use the Horiz Delay control to delay the acquisition relative to the B event.
6. Adjust the Ref Point and B Trig Level controls as necessary.
Behavior. When you select Trig on nth Event, the instrument looks for the trigger conditions specified by the A event. After the A
event arms the trigger system, the instrument looks for a specific number of B events before triggering. The instrument then
delays the acquisition for the amount of time specified by the Horiz Delay control. After the delay time has expired, the instrument
displays the data.
What do you want to do next?
Learn more about Horizontal Delay.
Learn more about other trigger types.
Return to the Trigger Setup control window.
Trigger on nth Event (Horizontal Delay Off).
Trigger on nth Event (B-Event Scan).
Trigger on nth Event (B-Event Scan and Horizontal Delay On).
Trigger setups
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series 453